TRANSMILLE : SOLUTIONS IN CALIBRATION Print PRINT SPECIFICATIONS
2006A :: LABORATORY REFERENCE
1 YEAR BEST ACCURACY SPECIFICATIONS

FUNCTION

SPECIFICATIONS MORE INFORMATION

Voltage DC


Range :
±0 to 1020V
Best Resolution : 0.1uV
Best 1 Year Accuracy : ±5ppm of setting


kV Amplifier available to extend
output to 10kV
Click here for details

Current DC


Range : ±0 to 20Amps
Best Resolution : 100pA
Best 1 Year Accuracy : ±20ppm of setting

Transconductance Amplifier
available to extend output to 100A
Click here for details

DC Resistance


Range : 0 to 1GOhm
Best 1 Year Accuracy : ±8ppm of setting

2 & 4 Wire Modes
Passive
decade resistors

Voltage AC


Range : 0 to 1020V
Best Resolution : 1uV
Frequency Range (Sine) : 10Hz to 100KHz
Best 1 Year Accuracy : ±0.01% of setting

kV Amplifier available to extend
output to 5kV
Click here for details

Current AC


Range : 0 to 20A
Best Resolution : 1nA
Frequency Range (Sine) : 10Hz to 10KHz
Best 1 Year Accuracy : ±0.03% of setting

Transconductance Amplifier
available to extend output to 70A
Click here for details

Capacitance


Range : 1nF to 100uF
Best 1 Year Accuracy : ±0.2% of setting

Passive decade capacitors for the best
compatibility with all bridges / meters
Inductance*


Range : 1mH to 10H
Best 1 Year Accuracy : ±0.5% of setting


Passive decade inductors for the best
compatibility with all bridges / meters

Thermocouple
Source*


Range : -210°C to 1800°C
Best 1 Year Accuracy : ±0.04°C of setting

7 Types Built-In
Additional types available using
Virtual Front Panel Software

PRT (RTD)
Source*


Range : -100°C TO 800°C (PT100)
Best 1 Year Accuracy : ±0.01% of setting

Additional RTD types available using
Virtual Front Panel Software
AC Power*


Range : 0.3W to 1MW
Phase Adjustment : 0 ± 180°
Frequency Range : 40 to 400Hz
Best 1 Year Accuracy : ±0.06% of setting

Incorporates Dynamic Digital Phase Compenstation (DDPC) to dynamically
control the phase angle between
voltage and current waveforms.
Eliminate errors caused by
capacitive or inductive loads
Phase*


Range : 0 to 359.9°
Best Resolution : 0.1°
Best Life Accuracy : 0.2°

Oscilloscope
Calibration*


Amplitude Range : 2mV to 50V / Div
Best 1 Year Accuracy : ±0.01% of setting

Timebase Range : 5s to 2ns / Div
Best 1 Year Accuracy : ±5ppm

Levelled Sweep Range : 5MHz to 600MHz
Best 1 Year Accuracy : 1db

50kHz Reference
Best 1 Year Accuracy (Frequency) : 30ppm
Best 1 Year Accuracy (Level) : 0.5%

Optical Tacho
Calibration*


Range : 240 to 60,000 RPM
Best 1 Year Accuracy : ±0.003%


Insulation Tester
Calibration*


Continuity Ohms Range : 1, 10 Ohm
Best 1 Year Accuracy : ±0.1% of setting


Insulation Ohms Range : 100k, 250k, 500k, 1M, 10M, 100M Ohm
Best 1 Year Accuracy : ±1% of setting

Insulation Voltage Range : 0 to 1000V
Best 1 Year Accuracy : ±5% of setting

Requires use of
Virtual Front Panel software (supplied)
or ProCal Calibration software
Only for use with a 2000 Series Calibrator
Power Supply
Calibration*


Voltage Range : 0 to 33V
Best 1 Year Accuracy : ±0.02% of setting



Current Range : 0 to 4A
Best 1 Year Accuracy : ±0.05% of setting


Requires use of
Virtual Front Panel software (supplied)
or ProCal Calibration software
Only for use with a 2000 Series Calibrator
GENERAL SPECIFICATIONS

Warm Up Time

Double the time since last used up to 20 minutes maximum

Standard Interface

RS232

Optional Interface

GPIB / USB

Temperature Performance

Storage : -5°C to +60°C
Operation : 0°C to +50°C
Calibration 18°C to 25°C

Temperature Coefficient

Specifications apply between 18°C and 25°C
Outside this range an allowance of 0.18 x 1 year specification per °C should be added

Relative Humidity

Operation <80% to 30°C, <70% to 40°C , <40% to 50°C
Storage : <95% (non-condensing)

Altitude

Operation : 3000m (10,00ft) Maximum
Transit : 12000m (40,000ft) Maximum

EMC & Safety

The calibrator line inlet plug must be earthed
See D.O.C. for full details

Low Analogue Isolation

100V

Line Power

Line voltage selectable : 110V / 230V
Line frequency : 50Hz to 60Hz
Line voltage variation : -6% +10%

Power Consumption

28 Watts (Standby)

Dimensions

Calibrator Only : 43cm x 46cm x 19mm
Calibrator in shipping box : 58cm x 56cm x 37cm
Calibrator in soft carry case : 49cm x 50cm x 19cm
Calibrator in hard transit case : 55cm x 56cm x 26cm

Weight

Calibrator Only : 12.5kg
Calibrator in shipping box : 15kg
Calibrator in soft carry case : 13.5
Calibrator in hard transit case : 22kg

* Denotes option
Due to continuous development specifications may be subject to change