| 1 YEAR BEST ACCURACY SPECIFICATIONS |
FUNCTION
|
SPECIFICATIONS
|
MORE INFORMATION |
Voltage DC
|
Range : ±0 to 1020V
Best Resolution : 0.1uV
Best 1 Year Accuracy : ±25ppm of setting
|
kV Amplifier available to
extend
output to
10kV
Click here for details |
Current DC |
Range : ±0 to 20Amps
Best Resolution : 100pA
Best 1 Year Accuracy : ±50ppm of setting
|
Transconductance Amplifier
available to extend
output to
100A
Click here for details |
DC Resistance |
Range : 0 to 1GOhm
Best 1 Year Accuracy : ±40ppm of setting
|
2 & 4 Wire Modes
Passive decade resistors |
Voltage AC |
Range : 0 to 1020V
Best Resolution : 1uV
Frequency Range (Sine) : 10Hz to 100KHz
Best 1 Year Accuracy : ±0.03% of setting
|
kV Amplifier available to extend
output to
5kV
Click here for details |
Current AC |
Range : 0 to 20A
Best Resolution : 1nA
Frequency Range (Sine) : 10Hz to 10KHz
Best 1 Year Accuracy : ±0.09% of setting
|
Transconductance Amplifier
available to extend output to
70A
Click here for details |
Capacitance |
Range : 1nF to 100uF
Best 1 Year Accuracy : ±0.25% of setting
|
Passive decade capacitors for the best
compatibility with all bridges / meters |
| Inductance* |
Range : 1mH to 10H
Best 1 Year Accuracy : ±0.5% of setting
|
Passive decade inductors for the best
compatibility with all bridges / meters |
Thermocouple
Source* |
Range : -210°C to 1800°C
Best 1 Year Accuracy : ±0.04°C of setting
|
7 Types Built-In
Additional types available using
Virtual Front Panel Software |
PRT (RTD)
Source* |
Range : -100°C TO 800°C (PT100)
Best 1 Year Accuracy : ±0.01% of setting
|
Additional RTD types available using
Virtual Front Panel Software |
| AC Power* |
Range : 0.3W to 1MW
Phase Adjustment : 0 ± 180°
Frequency Range : 40 to 400Hz
Best 1 Year Accuracy : ±0.11% of setting
|
Incorporates Dynamic Digital Phase Compenstation
(DDPC) to dynamically
control the phase angle
between
voltage and current waveforms.
Eliminate errors caused by
capacitive or inductive loads |
| Phase* |
Range : 0 to 359.9°
Best Resolution : 0.1°
Best Life Accuracy : 0.2°
|
Oscilloscope
Calibration*
|
Amplitude Range : 2mV to 50V / Div
Best 1 Year Accuracy : ±0.01% of setting
Timebase Range : 5s to 2ns / Div
Best 1 Year Accuracy : ±5ppm
Levelled Sweep Range : 5MHz to 600MHz
Best 1 Year Accuracy : 1db
50kHz Reference
Best 1 Year Accuracy (Frequency) : 30ppm
Best 1 Year Accuracy (Level) : ±0.5%
|
|
Optical Tacho
Calibration* |
Range : 240 to 60,000 RPM
Best 1 Year Accuracy : ±0.003%
|
|
Insulation Tester
Calibration* |
Continuity Ohms Range : 1, 10 Ohm
Best 1 Year Accuracy : ±0.1% of setting
Insulation Ohms Range : 100k, 250k, 500k, 1M, 10M, 100M Ohm
Best 1 Year Accuracy : ±1% of setting
Insulation Voltage Range : 0 to 1000V
Best 1 Year Accuracy : ±2% of setting
|
Requires use of
Virtual Front Panel software (supplied)
or ProCal Calibration software
Only for use with a 2000 Series Calibrator |
Power Supply
Calibration* |
Voltage Range : 0 to 33V
Best 1 Year Accuracy : ±0.02% of setting
Current Range : 0 to 4A
Best 1 Year Accuracy : ±0.05% of setting
|
Requires use of
Virtual Front Panel software (supplied)
or ProCal Calibration software
Only for use with a 2000 Series Calibrator |
| |
| GENERAL SPECIFICATIONS |
| |
|
Warm Up Time |
Double the time since last used up to 20 minutes maximum
|
Standard Interface |
RS232
|
Optional Interface |
GPIB / USB
|
Temperature Performance |
Storage : -5°C to +60°C
Operation : 0°C to +50°C
Calibration 17°C to 27°C
|
| Temperature Coefficient |
Specifications apply between 17°C and 27°C
Outside this range an allowance of 0.18 x 1 year specification per °C should be added
|
Relative Humidity |
Operation <80% to 30°C, <70% to 40°C , <40% to 50°C
Storage : <95% (non-condensing)
|
Altitude |
Operation : 3000m (10,00ft) Maximum
Transit : 12000m (40,000ft) Maximum
|
EMC & Safety |
The calibrator line inlet plug must be earthed
See D.O.C. for full details
|
Low Analogue Isolation |
100V
|
Line Power |
Line voltage selectable : 110V / 230V
Line frequency : 50Hz to 60Hz
Line voltage variation : -6% +10%
|
Power Consumption |
28 Watts (Standby)
|
Dimensions |
Calibrator Only : 43cm x 46cm x 19mm
Calibrator in shipping box : 58cm x 56cm x 37cm
Calibrator in soft carry case : 49cm x 50cm x 19cm
Calibrator in hard transit case
: 55cm x 56cm x 26cm
|
Weight |
Calibrator Only : 12.5kg
Calibrator in shipping box : 15kg
Calibrator in soft carry case : 13.5
Calibrator in hard transit case : 22kg
|
| |
|
* Denotes option
Due to continuous development specifications may be subject to change |
Back to top |