| 1 YEAR BEST ACCURACY SPECIFICATIONS |
FUNCTION
|
FUNCTION
|
MORE INFORMATION |
Voltage DC
|
Range : ±0 to 1020V
Best Resolution : 0.1uV
Best 1 Year Accuracy : ±50ppm of setting
|
kV Amplifier available to extend
output to 10kV
Click here for details |
Current DC |
Range : ±0 to 20Amps
Best Resolution : 100pA
Best 1 Year Accuracy : ±0.05% of setting
|
Transconductance Amplifier
available to extend output to 100A
Click here for details |
DC Resistance |
Range : 0 to 1GOhm
Best 1 Year Accuracy : ±0.005% of setting
|
2 Wire Mode
Passive decade resistors |
Voltage AC |
Range : 0 to 1020V
Best Resolution : 1uV
Frequency Range (Sine) : 40Hz to 10KHz
Best 1 Year Accuracy : ±0.07% of setting
|
kV Amplifier available to extend
output to 5kV
Click here for details |
Current AC |
Range : 0 to 20A
Best Resolution : 1nA
Frequency Range (Sine) : 40Hz to 500Hz
Best 1 Year Accuracy : ±0.1% of setting
|
Transconductance Amplifier
available to extend output to 70A
Click here for details |
Capacitance |
Range : 10nF & 1uF
Best 1 Year Accuracy : ±0.4% of setting
|
Passive decade capacitors for the best
compatibility with all bridges / meters |
Thermocouple
Source |
Range : -210°C to 1800°C
Best 1 Year Accuracy : ±0.04°C of setting
|
7 Types Built-In
Additional types available using
Virtual Front Panel Software |
AC Power*
(Part of combined Scope/Power option) |
Range : 0.3W to 1MW
Phase Adjustment : 0 ± 180° (0.1° steps)
Frequency Range : 40 to 400Hz
Best 1 Year Accuracy : ±0.11% of setting
|
Incorporates Dynamic Digital Phase Compenstation (DDPC) to dynamically
control the phase angle between
voltage and current waveforms.
Eliminate errors caused by
capacitive or inductive loads |
Phase*
(Part of combined Scope/Power option |
Range : 0 to 359.9°
Best Resolution : 0.1°
Best Life Accuracy : 0.2°
|
Oscilloscope*
Calibration
(Part of combined Scope/Power option
|
Amplitude Range : 2mV to 50V / Div
Best 1 Year Accuracy : ±0.01% of setting
Timebase Range : 5s to 2ns / Div
Best 1 Year Accuracy : ±5ppm
Levelled Sweep Range : 5MHz to 250MHz
Best 1 Year Accuracy : 1db
50kHz Reference
Best 1 Year Accuracy (Frequency) : 30ppm
Best 1 Year Accuracy (Level) : ±0.5%
|
|
| |
| GENERAL SPECIFICATIONS |
| |
|
Warm Up Time |
Double the time since last used up to 20 minutes maximum
|
Standard Interface |
RS232
|
Optional Interface |
GPIB / USB
|
Temperature Performance |
Storage : -5°C to +60°C
Operation : 0°C to +50°C
Calibration 17°C to 30 °C
|
| Temperature Coefficient |
Specifications apply between 17°C and 30 °C
Outside this range an allowance of 0.18 x 1 year specification per °C should be added
|
Relative Humidity |
Operation <80% to 30°C, <70% to 40°C , <40% to 50°C
Storage : <95% (non-condensing)
|
Altitude |
Operation : 3000m (10,00ft) Maximum
Transit : 12000m (40,000ft) Maximum
|
EMC & Safety |
The calibrator line inlet plug must be earthed
See D.O.C. for full details
|
Low Analogue Isolation |
100V
|
Line Power |
Line voltage selectable : 110V / 230V
Line frequency : 50Hz to 60Hz
Line voltage variation : -6% +10%
|
Power Consumption |
28 Watts (Standby)
|
Dimensions |
Calibrator Only : 43cm x 46cm x 19mm
Calibrator in shipping box : 58cm x 56cm x 37cm
Calibrator in soft carry case : 49cm x 50cm x 19cm
Calibrator in hard transit case : 55cm x 56cm x 26cm
|
Weight |
Calibrator Only : 12.5kg
Calibrator in shipping box : 15kg
Calibrator in soft carry case : 13.5
Calibrator in hard transit case : 22kg
|
| |
|
* Denotes option
Due to continuous development specifications may be subject to change |
Back to top |