Checking of Insulation Testers
3 high value resistors (1M Ohms, 9.9M Ohms & 99M Ohms) for testing the resistance measurement function.
LED indication of test voltage at 100V 250V 500V 1000V
3 low value resistors ( 0.5 Ohms, 2 Ohms & 10 Ohms) for testing the continuity function.
AC line voltage output for testing voltage measurement

Checking of RCD Testers
Dedicated 13A socket provides functional check of RCD tester operation
3 selectable ranges : 10mA@150ms 30mA@150ms 150mA@30ms
Indicators for Ready Tripped Overcurrent

Checking of LOOP Testers
A functional check of LOOP Testers, allowing direct measurement of supplied loop impedance and supplied loop impedance + 1 Ohm via a dedicated 13A socket.

Wiring Check
When connected, the 2080 checkbox will indicate voltage between the following wires *
P-E (Phase to Earth) P-N (Phase to Neutral) N-E (Neutral to Earth)
* Does not detect Neutral to Earth reversal
Certification
Calibration Certificates traceable to UK National Standards or UKAS are available.
|