Pressing the CLASS 1 button sets the 6080 to automatically sequence through a visual check, fuse check, earth bond resistance, insulation resistance and leakage current test for each appliance.
The CLASS 2 button performs the same sequence without the earth bond test.
Pass or fail is clearly shown using bright red/green LED bar displays for each test - this makes reading and understanding the results easy for all levels of users. Pass / fail limits scale automatically with the earth bond current used, the lead length and the test type. Readings are also displayed on a digital LCD.
6080 :: ADDITIONAL FUNCTIONALITY
Expanding on the strengths of the 5080 PAT , the 6080 adds dual voltage capabilities for complete 110V / 240V compatibility with separate outputs for each voltage.
The IEC / extension lead check function leads to be verified independently. The self check function allows the 6080 to be verified using its own built in test facility - simply connect to the self check outputs to confirm correct PAT operation
Fuse GO / NO GO function allows fuse continuity to be checked with LED indication of result.
The 6080 includes larger battery capacity to provide longer test times between charging.
PC CONNECTION :: DATA DOWNLOAD & REPORTS
All tests performed are stored in the 5080 memory together with a unique test number and date.
Downloading the results from the 5080 is a snap using the PC connection kit (option).
One click is all that is required to fetch the data and display a report of all your work done.
Easy function selection - Download, Upload, Edit Appliances, Print Reports or Print Labels
'1 CLICK' Download - Simply click Download Data on PC.
Press any button on the PAT to begin download
Easy Upload - Simply select appliances and click Upload Data on PC.
Press any button on the PAT to begin upload
Easy Report Printing - enter customer/location details
and tested by name then click PRINT